Isotopic internal standards are increasingly frequent in LC-MS analysis to control biological matrix effects in the quantitation of immunosuppressant drugs, such as everolimus (RAD001). Here we present the evaluation of a LC-MS method, exploiting [13C2D4]RAD001 as internal standard, for preclinical determination of RAD001 in mice brain tissue. Samples were purified by solid phase extraction. Brain and blood were collected from vehicle-treated and RAD001-treated mice. The QTOF MS detector was set to select RAD001 ammonium adducts (m/z 975.6152) and [13C2D4]RAD001 (m/z 981.6481). Two different UHPLC columns were preliminarily tested. The method showed linear behavior between 4-100 ng/mL (r2=0.99943) and linearity was preserved in the presence of blood (r2=0.99107) and brain (r2=0.99098) matrix components. Intra-day and inter-day precision (3%-19%) and accuracy (82%-109%) were comparable between standards and spiked blood and brain samples. As resulting from recovery comparison (82%-98%), [13C2D4]RAD001 compensated ion suppression phenomena maintaining method performance over a wide range of consecutive analytical runs. The comparison with a HPLC-UV method showed reliability of the method with good correlation between blood (r2=0.94319) and brain (r2=0.97773) samples and acceptable biases (<15%). This validation suggests that the investigated method could be useful for the preclinical monitoring of RAD001 brain therapeutic concentrations in animal models.

Evaluation of a LC-MS method for everolimus preclinical determination in brain by using [13C2D4]RAD001 internal standard

GIOVAGNOLI, Stefano
;
MAGINI, Alessandro;POLCHI, ALICE;TANCINI, Brunella;EMILIANI, Carla;
2015

Abstract

Isotopic internal standards are increasingly frequent in LC-MS analysis to control biological matrix effects in the quantitation of immunosuppressant drugs, such as everolimus (RAD001). Here we present the evaluation of a LC-MS method, exploiting [13C2D4]RAD001 as internal standard, for preclinical determination of RAD001 in mice brain tissue. Samples were purified by solid phase extraction. Brain and blood were collected from vehicle-treated and RAD001-treated mice. The QTOF MS detector was set to select RAD001 ammonium adducts (m/z 975.6152) and [13C2D4]RAD001 (m/z 981.6481). Two different UHPLC columns were preliminarily tested. The method showed linear behavior between 4-100 ng/mL (r2=0.99943) and linearity was preserved in the presence of blood (r2=0.99107) and brain (r2=0.99098) matrix components. Intra-day and inter-day precision (3%-19%) and accuracy (82%-109%) were comparable between standards and spiked blood and brain samples. As resulting from recovery comparison (82%-98%), [13C2D4]RAD001 compensated ion suppression phenomena maintaining method performance over a wide range of consecutive analytical runs. The comparison with a HPLC-UV method showed reliability of the method with good correlation between blood (r2=0.94319) and brain (r2=0.97773) samples and acceptable biases (<15%). This validation suggests that the investigated method could be useful for the preclinical monitoring of RAD001 brain therapeutic concentrations in animal models.
2015
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11391/1314899
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact