A 55-year-old male with a history of severe organic mitral regurgitation treated with surgical mitral valve (MV) repair was referred for a transcatheter MV replacement due to recurrent regurgitation. After the release of the first transcatheter MV, a severe paravalvular leak coming from the lateral side was observed. To promptly tackle this issue, a second valve with further postdilation was successfully implanted and the paravalvular leak disappeared. This case highlights the feasibility of implanting a second valve in case of severe paravalvular leaks after MV-in-ring procedures due to device malapposition.

Valve-in-Valve-in-Ring: A Bailout Strategy to Tackle Paravalvular Leaks due to Device Malapposition

Fortuni F;
2022

Abstract

A 55-year-old male with a history of severe organic mitral regurgitation treated with surgical mitral valve (MV) repair was referred for a transcatheter MV replacement due to recurrent regurgitation. After the release of the first transcatheter MV, a severe paravalvular leak coming from the lateral side was observed. To promptly tackle this issue, a second valve with further postdilation was successfully implanted and the paravalvular leak disappeared. This case highlights the feasibility of implanting a second valve in case of severe paravalvular leaks after MV-in-ring procedures due to device malapposition.
2022
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11391/1589198
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact