MRI is a widespread and greatly helpful diagnostic tool, yet its use on cochlear implant patients is restricted by the presence of an inner magnet. We report on a case of magnet dislodgment after 1.5T MRI in a 31-year-old female with a Hi-Res 90K cochlear implant. In this case, it was possible to implement an alternative, totally noninvasive approach based on an external manual repositioning rather than a surgical procedure of the displaced magnet. Laryngoscope, 2013 Copyright © 2012 The American Laryngological, Rhinological, and Otological Society, Inc.

Noninvasive management of cochlear implant's inner magnet displacement after magnetic resonance

De Corso E.;
2013

Abstract

MRI is a widespread and greatly helpful diagnostic tool, yet its use on cochlear implant patients is restricted by the presence of an inner magnet. We report on a case of magnet dislodgment after 1.5T MRI in a 31-year-old female with a Hi-Res 90K cochlear implant. In this case, it was possible to implement an alternative, totally noninvasive approach based on an external manual repositioning rather than a surgical procedure of the displaced magnet. Laryngoscope, 2013 Copyright © 2012 The American Laryngological, Rhinological, and Otological Society, Inc.
2013
File in questo prodotto: