Interventional radiologists and staff members are frequently exposed to protracted and fractionated low doses of ionizing radiation due to diffused X-ray radiation. The authors propose a novel approach to monitor on line staff during their interventions by using a device based on an Active Pixel Sensor developed for tracking applications. Two different photodiode configurations have been tested in standard Interventional Radiology working conditions. Both options have demonstrated the capability to measure the photon flux and the energy flux to a sufficient degree of uncertainty.

An active pixel sensor to detect diffused X-ray during Interventional Radiology procedure

BIASINI, Maurizio;CONTI, ELIA;FANO', Livio;PASSERI, Daniele;PLACIDI, Pisana
2012

Abstract

Interventional radiologists and staff members are frequently exposed to protracted and fractionated low doses of ionizing radiation due to diffused X-ray radiation. The authors propose a novel approach to monitor on line staff during their interventions by using a device based on an Active Pixel Sensor developed for tracking applications. Two different photodiode configurations have been tested in standard Interventional Radiology working conditions. Both options have demonstrated the capability to measure the photon flux and the energy flux to a sufficient degree of uncertainty.
2012
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11391/911429
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact